A Non-Magnetic RF Balun Designed at 128 MHz Centre frequency for 3 T MRI Scanners

Document Type

Conference Proceeding

Publication Date

8-30-2020

Journal

Canadian Conference on Electrical and Computer Engineering

Volume

2020-August

URL with Digital Object Identifier

10.1109/CCECE47787.2020.9255767

Abstract

In this work, a non-magnetic RF balun is implemented for 3 T magnetic resonance imaging (MRI) scanners operating at 128 MHz to transform a balanced input signal from a dipole or loop antenna into an unbalanced output signal. It is fabricated on a low-cost, copper cladded four-layer printed circuit board (PCB), FR4 with a thickness of 1.57 mm and a copper thickness of 35 ?m, with overall footprint of 11.6 mm × 12.2 mm. A comparison among the ADS RF momentum simulations and the measured results indicates a good agreement with the measured insertion and return losses of better than -1 dB and -13 dB, respectively, in a 50 ? termination setting.