Ultra-low frequency magnetic field probe for field measurements in gradient coils during medical device testing

Document Type

Article

Publication Date

1-1-2019

Journal

Electronics Letters

Volume

55

Issue

11

First Page

634

Last Page

636

URL with Digital Object Identifier

10.1049/el.2019.0204

Abstract

© The Institution of Engineering and Technology 2019. In this Letter ultra-low frequency probes for measuring time-varying magnetic field produced in gradient coils were designed, fabricated, and tested. A 1 cm radius induction loop antenna followed by an instrumentational amplifier with a gain of 40 dB and a cutoff frequency of 30 KHz was designed to measure the magnetic field. The probe was fabricated on a single-sided PCB, FR4 of thickness 1.57 mm and copper of thickness 35 μm was employed in a dB/dt exposure system within a saline phantom using 3D printed waterproofed cover. Simulation and measurement results of the magnetic field were in good agreements.