Application and experimental validation of an integral method for simulation of gradient-induced eddy currents on conducting surfaces during magnetic resonance imaging.

Document Type

Article

Publication Date

6-21-2013

Journal

Physics in medicine and biology

Volume

58

Issue

12

First Page

4367

Last Page

4379

URL with Digital Object Identifier

10.1088/0031-9155/58/12/4367

Abstract

The time-varying magnetic fields created by the gradient coils in magnetic resonance imaging can produce negative effects on image quality and the system itself. Additionally, they can be a limiting factor to the introduction of non-MR devices such as cardiac pacemakers, orthopedic implants, and surgical robotics. The ability to model the induced currents produced by the switching gradient fields is key to developing methods for reducing these unwanted interactions. In this work, a framework for the calculation of induced currents on conducting surface geometries is summarized. This procedure is then compared to two separate experiments: (1) the analysis of the decay of currents induced upon a conducting cylinder by an insert gradient set within a head only 7 T MR scanner; and (2) analysis of the heat deposited into a small conductor by a uniform switching magnetic field at multiple frequencies and two distinct conductor thicknesses. The method was shown to allow the accurate modeling of the induced time-varying field decay in the first case, and was able to provide accurate estimation of the rise in temperature in the second experiment to within 30% when the skin depth was greater than or equal to the thickness of the conductor.

Notes

This article is paywalled; please seek access through your institution's library

Find in your library

Share

COinS