Bone and Joint Institute
Wear performance of cobalt chromium, ceramic, and oxidized zirconium on highly crosslinked polyethylene at mid-term follow-up
Document Type
Article
Publication Date
6-1-2018
Journal
Journal of Orthopaedics
Volume
15
Issue
2
First Page
620
Last Page
623
URL with Digital Object Identifier
10.1016/j.jor.2018.05.018
Abstract
© 2018 Prof. PK Surendran Memorial Education Foundation Patients with THA using a ceramic head and using an OxZr head were each matched to patients with a CoCr head. Mean implantation time was 5 years. There was no difference in steady state wear rate between the ceramic (0.066 ± 0.050 mm/year) and CoCr match groups (0.052 ± 0.041 mm/year), or between the OxZr (0.022 ± 0.029 mm/year) and CoCr match groups (0.048 ± 0.071 mm/year). Follow-up into the second decade will be necessary before any changes in THA wear rate from using ceramic or OxZr bearings may be appreciated clinically with available imaging techniques.