Characterization of various tissue mimicking materials for medical ultrasound imaging
Document Type
Conference Proceeding
Publication Date
1-1-2016
Journal
Progress in Biomedical Optics and Imaging - Proceedings of SPIE
Volume
9783
URL with Digital Object Identifier
10.1117/12.2218160
Abstract
© 2016 SPIE. Tissue mimicking materials are physical constructs exhibiting certain desired properties, which are used in machine calibration, medical imaging research, surgical planning, training, and simulation. For medical ultrasound, those specific properties include acoustic propagation speed and attenuation coefficient over the diagnostic frequency range. We investigated the acoustic characteristics of polyvinyl chloride (PVC) plastisol, polydimethylsiloxane (PDMS), and isopropanol using a time-of-light technique, where a pulse was passed through a sample of known thickness contained in a water bath. The propagation speed in PVC is approximately 1400ms-1 depending on the exact chemical composition, with the attenuation coefficient ranging from 0:35 dB cm-1 at 1MHz to 10:57 dB cm-1 at 9 MHz. The propagation speed in PDMS is in the range of 1100ms-1, with an attenuation coefficient of 1:28 dB cm-1 at 1MHz to 21:22 dB cm-1 at 9 MHz. At room temperature (22 °C), a mixture of water-isopropanol (7:25% isopropanol by volume) exhibits a propagation speed of 1540ms-1, making it an excellent and inexpensive tissue-mimicking liquid for medical ultrasound imaging.