Title

Assessment of the best flow model to characterize diffuse correlation spectroscopy data acquired directly on the brain

Document Type

Article

Publication Date

10-7-2015

Journal

Biomedical Optics Express

First Page

4288

Last Page

4301

URL with Digital Object Identifier

10.1364/BOE.6.004288

Abstract

© 2015 Optical Society of America. Diffuse correlation spectroscopy (DCS) is a non-invasive optical technique capable of monitoring tissue perfusion. The normalized temporal intensity autocorrelation function generated by DCS is typically characterized by assuming that the movement of erythrocytes can be modeled as a Brownian diffusion-like process instead of by the expected random flow model. Recently, a hybrid model, referred to as the hydrodynamic diffusion model, was proposed, which combines the random and Brownian flow models. The purpose of this study was to investigate the best model to describe autocorrelation functions acquired directly on the brain in order to avoid confounding effects of extracerebral tissues. Data were acquired from 11 pigs during normocapnia and hypocapnia, and flow changes were verified by computed tomography perfusion (CTP). The hydrodynamic diffusion model was found to provide the best fit to the autocorrelation functions; however, no significant difference for relative flow changes measured by the Brownian and hydrodynamic diffusion models was observed.

This document is currently not available here.